Optical and radiometric models of the NOMAD instrument part I: the UVIS channel
Metadata
Afficher la notice complèteAuthor
Vandaele, Ann C.
Willame, Yannick
Depiesse, Cédric
Thomas, Ian R.
Robert, Séverine
Bolsée, David
Patel, Manish R.
Mason, Jon P.
Leese, Mark
Lesschaeve, Stefan
Antoine, Philippe
Daerden, Frank
Delanoye, Sofie
Drummond, Rachel
Neefs, Eddy
Ristic, Bojan
Lopez-Moreno, José-Juan
Bellucci, Giancarlo
NOMAD Team, NOMAD Team
Date
2015-11-09Citation
Vandaele, A.C. Y. Willame, C. Depiesse, I. R. Thomas, S. Robert, D. Bolsée, M. R. Patel, J. P. Mason, M. Leese, S. Lesschaeve, P. Antoine, F. Daerden, S. Delanoye, R. Drummond, E. Neefs, B. Ristic, J. J. Lopez-Moreno, G. Bellucci, and the NOMAD Team. "Optical and radiometric models of the NOMAD instrument Part I: the UVIS channel." Optics Express 23 (2015): 30028–30042 (2015). DOI: 10.1364/OE.23.030028.
Abstract
The NOMAD instrument has been designed to best fulfil the science objectives of the ExoMars Trace Gas Orbiter mission that will be launched in 2016. The instrument is a combination of three channels that cover the UV, visible and IR spectral ranges and can perform solar occultation, nadir and limb observations. In this series of two papers, we
present the optical models representing the three channels of the instrument and use them to determine signal to noise levels for different observation modes and Martian conditions. In this first part, we focus on the UVIS channel, which will sound the Martian atmosphere using nadir and solar occultation viewing modes, covering the 200-650nm spectral range. High SNR levels (>1000) can easily be reached for wavelengths higher than 300nm both in solar occultation and nadir modes when considering binning.
Below 300nm SNR are lower primarily because of the lower signal and the impact of atmospheric absorption.